| Model Number |
CQ75104 |
| Medical Device Problem Codes |
Break (1069); Inflation Problem (1310); Material Twisted/Bent (2981)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As the lot number for the device was provided, a lot history review was performed.The sample was returned to the manufacturer for inspection/evaluation.The investigation is confirmed for inflation issue, break and twisted inner catheter.A definitive root cause for the reported event could not be determined.The device is labeled for single use.
|
| |
|
Event or Problem Description
|
|
This report summarizes one malfunction.A review of the reported information indicated that model cq75104 pta balloon dilatation catheter allegedly experienced inflation problem, break and material twisted.This information was received from one source.The event involved a patient with no known impact to the patient.The (b)(6) year old female patient was (b)(6) lbs.
|
| |
|