| Model Number |
MC2-5590R |
| Medical Device Problem Codes |
Migration or Expulsion of Device (1395); Migration (4003)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
No product has been returned for evaluation nor were radiographs provided to confirm the alleged event.No root cause can be confirmed at this time.Device in-situ.
|
| |
|
Event or Problem Description
|
|
Information was received that prior to the first distraction, x -ray images revealed loosening of the end cap.A revision procedure is not planned at this time.There was no patient injury reported.As per the surgeon, the initial distraction session went well.
|
| |
|
Additional Manufacturer Narrative
|
|
Even though no product has been returned, fsca investigation was performed and variation in the torque applied to the threaded cap during the assembly process was identified as the root cause of the failure.Per the manufacturing instructions, the threaded cap must be tightened to 40 in-lbs.Although all operators followed the assembly procedure, and the torque wrench indicated 40 in-lbs was applied, the manner of using the torque wrench resulted in variances in applied torque.The manner in which the operator handles the torque wrench may impact the actual torque applied to the end cap, thereby creating a false positive that the specified torque has been applied.If the specified torque is not applied, the effectiveness of the cap tightening process may be compromised.
|
| |
|