Medtronic received information that it was not possible to collect the flip.During the microcatheter withdrawal, the flip was stuck in the distal part of the pipeline causing it to be displaced.No other information was provided besides the treatment being for an aneurysm.Additional information received reporting that the original mention of the "flip" was in reference to the ptfe sleeves.As the pipeline was displaced, it did not fully cover the neck of the aneurysm so a second pipeline was also implanted was also implanted for coverage which was successful.The surgeon reported the sleeves were ultimately removed together with the catheter and the devices were discarded by the hospital.As the issue was known during surgery, it was addressed during the procedure without injury to the patient.
|