• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS, INCORPORATED MAGEC SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS, INCORPORATED MAGEC SYSTEM; GROWING ROD SYSTEM - MAGNETIC ACTUATION Back to Search Results
Model Number MC2-5070S
Medical Device Problem Codes Defective Component (2292); Migration (4003)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 06/02/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device has not yet been returned.Root cause is not available at this time.
 
Event or Problem Description
Information was received that a revision procedure was performed on (b)(6) 2020.As per the reporter, x-ray images revealed that the end cap had disengaged.There was no patient injury reported.
 
Additional Manufacturer Narrative
Even though no product has been returned, fsca investigation was performed and variation in the torque applied to the threaded cap during the assembly process was identified as the root cause of the failure.Per the manufacturing instructions, the threaded cap must be tightened to 40 in-lbs.Although all operators followed the assembly procedure, and the torque wrench indicated 40 in-lbs was applied, the manner of using the torque wrench resulted in variances in applied torque.The manner in which the operator handles the torque wrench may impact the actual torque applied to the end cap, thereby creating a false positive that the specified torque has been applied.If the specified torque is not applied, the effectiveness of the cap tightening process may be compromised.
 
Additional Manufacturer Narrative
The product was returned and the reported failure mode was confirmed.Visual inspection of the returned product confirmed the end cap was unthreaded from the housing tube.The returned rod was observed to be partially distracted.X-ray imaging inspection was also performed; end cap disengagement was clear.Per reported failure, functional testing was not applicable.Review of the device history records indicated the rod was functionally tested.The issue recorded in this complaint is a known issue and has been investigated.
 
Event or Problem Description
See updated information in h2, h3, h6, h10.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAGEC SYSTEM
Common Device Name
GROWING ROD SYSTEM - MAGNETIC ACTUATION
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INCORPORATED
101 enterprise dr, suite 100
aliso viejo CA 92656
MDR Report Key10292225
Report Number3006179046-2020-00343
Device Sequence Number3376799
Product Code PGN
UDI-Device Identifier00887517834027
UDI-Public887517834027
Combination Product (Y/N)N
Initial Reporter StateAL
Initial Reporter CountryUS
PMA/510(K) Number
K171791
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Remedial Action Notification,Recall
Type of Report Initial,Followup,Followup
Report Date (Section B) 02/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberMC2-5070S
Device Lot Number8041114AAA
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/17/2020
Supplement Date Received by Manufacturer06/17/2020
01/13/2021
Initial Report FDA Received Date07/17/2020
Supplement Report FDA Received Date09/30/2020
02/01/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
Z-1898-2020
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age10 YR
Patient Weight23
-
-