| Model Number |
Fibula Cutting Guide with Implant Placement |
| Medical Device Problem Code |
Patient-Device Incompatibility (2682)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
06/25/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The dental implant placement on the fibula cutting guide with implant placement was placed on the anterior surface instead of the posterior surface.The issue was identified prior to surgery and the implant placement feature on the guide was not used.The issue caused no harm to the patient and did not cause a delay in surgery.
|
| |
|
Event or Problem Description
|
|
The dental implant placement on the fibula cutting guide with implant placement was placed on the anterior surface instead of the posterior surface.The issue was identified prior to surgery and the implant placement feature on the guide was not used.The issue caused no harm to the patient and did not cause a delay in surgery.
|
| |
|