| Model Number |
3224 |
| Medical Device Problem Code |
High impedance (1291)
|
| Health Effect - Clinical Code |
Inadequate Pain Relief (2388)
|
| Date of Event |
06/29/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The patient's weight was unable to be obtained.The results/method and conclusion codes along with the investigation results will be provided in the final report.
|
| |
|
Event or Problem Description
|
|
Device 1 of 2; reference mfr.Report#: 1627487-2020-22955.It was reported the high impedance was found on both of the patient's lead contacts.As a result, both of the patient's leads were explanted and replaced to address their issue.
|
| |
|
Event or Problem Description
|
|
Device 1 of 2 reference mfr.Report#: 1627487-2020-22955.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported event of high impedance was not confirmed.As received the paddle length was complete but cut into two pieces.It passed continuity testing on all channels.No root cause was identified for the reported event.
|
| |
|