| Model Number |
AIRLIFE¿ MANUAL RESUSCITATOR VARIABLE VOLUME OXYGEN RESERVOIR TUBING, ADULT MASK |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/18/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
At this time, the suspect device has not been returned for evaluation.Therefore, root cause has not been determined yet.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information becomes available.
|
| |
|
Event or Problem Description
|
|
The customer reported that the tube of airlife adult manual resuscitator, variable volume oxygen reservoir tubing, adult mask comes out and cannot be reattached.The issue occurred during patient-use and the customer confirmed that there was no patient harm associated with this reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Result of investigation: the device history record (dhr) review did not show any deviations to manufacturer specification.For functionality failure, it is imperative to receive a sample to duplicate the failure as described.
|
| |
|
Additional Manufacturer Narrative
|
|
Result of investigation: 15 open samples of the p/n 2k8017 without lot number were received for the investigation, samples were evaluated and found 14 samples defectives, the oxygen line (tubing) is disconnected from threaded o2 housing and also found some threaded o2 housing broken, these samples were handled by the customer and it is possible that they damaged the components.It was also observed in the photograph that the oxygen line (tubing) is disconnected from threaded o2 housing.Vyaire found that if mix adhesive used on the assemble with alcohol provoke that tubing could disconnect.The alcohol affect the effective of the adhesive.
|
| |
|