| Model Number |
MMT-523NAP |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/13/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that insulin pump had motor sensor test failure.Customer was able to clear the alarm and unable to complete rewind.Customer was performed displacement test and it was failed.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with detached end cap.Unable to perform functional testing including the displacement, rewind, basic occlusion, occlusion, prime/a33 and excessive no delivery test or verify a35 alarm due to detached end cap.
|
| |
|