| Model Number |
MMT-523NAB |
| Medical Device Problem Codes |
Connection Problem (2900); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/22/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the reservoir keeps falling out of the pump.The customer stated that, the retainer ring on the pump is missing.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with reservoir tube lip completely broken off and missing the black o ring noted.The test reservoir does not stay locked in place due to reservoir tube lip broken off.Unable to perform displacement test due to reservoir lip broken.
|
| |
|
Search Alerts/Recalls
|