| Model Number |
MMT-7020A |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/12/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic the user reported inaccurate sensor readings that triggered a threshold suspend.The customer¿s blood glucose was 120 mg/dl and sensor glucose was 75 mg/dl at the time of the event, the difference is outside the acceptable range.Suspend on low limit in sensor settings was at 75 mg/dl.No harm requiring medical intervention was reported.The sensor will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Inspected one opened and used sensor and performed continuity resistance test, sensor passed per specification.Then performed a bicarbonate buffer test, and sensor passed per specifications with accurate readings.(b)(4).
|
| |
|