• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SOFRADIM PRODUCTION SAS PARIETEX; MESH, SURGICAL, POLYMERIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SOFRADIM PRODUCTION SAS PARIETEX; MESH, SURGICAL, POLYMERIC Back to Search Results
Model Number PCO6VP
Medical Device Problem Codes Migration or Expulsion of Device (1395); Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Purulent Discharge (1812); Edema (1820); Foreign Body Reaction (1868); Inflammation (1932); Pain (1994); Swelling (2091); Discomfort (2330); Injury (2348); No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
The patient¿s attorney alleged a deficiency against the device.The product was used for therapeutic treatment of an umbilical hernia.It was reported that after the implant, the patient experienced hypervascularity, wadded up mesh, inflammation, pain, edema, swelling, purulent material, tenderness, and abnormal foreign body reaction.Post-operative patient treatment included revision surgery, hernia repair with new mesh, and removal of mesh.
 
Additional Manufacturer Narrative
Information received shows that this product event is a duplicate of another issue reported under regulatory report #9615742-2019-01460.This file will be closed and all further updates processed under the above-referenced report.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARIETEX
Common Device Name
MESH, SURGICAL, POLYMERIC
Manufacturer (Section D)
SOFRADIM PRODUCTION SAS
116 avenue du formans
trevoux 01600
FR  01600
MDR Report Key10295460
Report Number9615742-2020-01668
Device Sequence Number17015395
Product Code FTL
UDI-Device Identifier10884521165717
UDI-Public10884521165717
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K120506
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2018
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 07/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2019
Device Model NumberPCO6VP
Device Catalogue NumberPCO6VP
Device Lot NumberPOC8040
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/01/2020
Supplement Date Received by Manufacturer07/01/2020
Initial Report FDA Received Date07/20/2020
Supplement Report FDA Received Date07/28/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-