• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL - NEUROMODULATION LAMITRODE S-8 LEAD KIT, 60CM LENGTH; SCS PADDLE LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL - NEUROMODULATION LAMITRODE S-8 LEAD KIT, 60CM LENGTH; SCS PADDLE LEAD Back to Search Results
Model Number 3286
Medical Device Problem Code Migration (4003)
Health Effect - Clinical Code Inadequate Pain Relief (2388)
Date of Event 06/26/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Event date is estimated.The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event or Problem Description
Related manufacturer reference numbers: 1627487-2020-22792, 3006705815-2020-02779.It was reported that the patient sustained a fall which lead to lead migration.In turn, the patient underwent surgical intervention wherein the entire scs system was explanted and replaced.Post-operatively, stimulation therapy was restored.
 
Additional Manufacturer Narrative
During processing of this complaint, attempts were made to obtain complete patient information.The reported event of lead migration cannot be confirmed with product testing alone.Return paddle lead was damage beyond functional testing.No root cause was identified for the reported issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LAMITRODE S-8 LEAD KIT, 60CM LENGTH
Common Device Name
SCS PADDLE LEAD
Manufacturer (Section D)
ST. JUDE MEDICAL - NEUROMODULATION
6901 preston rd
plano TX 75024
MDR Report Key10295519
Report Number1627487-2020-22935
Device Sequence Number14389843
Product Code LGW
UDI-Device Identifier05414734406291
UDI-Public05414734406291
Combination Product (Y/N)N
Initial Reporter StatePR
Initial Reporter CountryUS
PMA/510(K) Number
P010032
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2016
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,consum
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date01/31/2017
Device Model Number3286
Device Catalogue Number3286
Device Lot Number4995216
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/20/2020
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/26/2020
Supplement Date Received by Manufacturer08/26/2020
Initial Report FDA Received Date07/20/2020
Supplement Report FDA Received Date09/02/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
PERCUTANEOUS LEAD; SCS IPG; PERCUTANEOUS LEAD; SCS IPG
Outcome Attributed to Adverse Event Other;
-
-