• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH MOUNTAINEER OCT SPINAL SYSTEM MINI POLYAXIAL BONE SCREW 3.5 X 3.5 X 32MM; ORTHOSIS, CERVICAL PEDICLE SCREW SPINAL FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH MOUNTAINEER OCT SPINAL SYSTEM MINI POLYAXIAL BONE SCREW 3.5 X 3.5 X 32MM; ORTHOSIS, CERVICAL PEDICLE SCREW SPINAL FIXATION Back to Search Results
Model Number 188320332
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Failure of Implant (1924); Injury (2348)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2020 the patient had a fall and fell face first resulting in the breaking of mountaineer screws at c1 and c2 midshaft.A revision surgery is pending the mri and ct results.Concomitant devices: unknown setscrews (part# unknown, lot# unknown, quantity unknown); unknown rods (part# unknown, lot# unknown, quantity unknown); unknown occipital plate (part# unknown, lot# unknown, quantity unknown).This report is for one (1) mountaineer oct spinal system mini polyaxial bone screw 3.5 x 3.5 x 32mm.This is report 1 of 4 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h6: customer quality investigation: the complaint device was not received for investigation.The following investigation is based on the image(s) provided.The image(s) was reviewed, and the complaint condition could be confirmed as the x-ray image clearly displays that the screw was broken into 2+ pieces.Since the device was not returned, a dimensional inspection and a functional test were not able to be performed.A manufacturing record evaluation could not be performed as the lot number could not be determined from the image provided.A definitive assignable root cause could not be determined based on the provided information.During the investigation no product design issues or discrepancies were observed (based on the images) that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.H11 corrected data: g1: manufacturer's information.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOUNTAINEER OCT SPINAL SYSTEM MINI POLYAXIAL BONE SCREW 3.5 X 3.5 X 32MM
Common Device Name
ORTHOSIS, CERVICAL PEDICLE SCREW SPINAL FIXATION
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
MDR Report Key10295795
Report Number1526439-2020-01354
Device Sequence Number11362424
Product Code NKG
UDI-Device Identifier10705034158143
UDI-Public10705034158143
Combination Product (Y/N)N
Initial Reporter StateOK
Initial Reporter CountryUS
PMA/510(K) Number
K041203
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number188320332
Device Catalogue Number188320332
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/01/2020
Supplement Date Received by Manufacturer07/31/2020
Initial Report FDA Received Date07/20/2020
Supplement Report FDA Received Date07/31/2020
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNKNOWN LOCKING/SET SCREWS; UNKNOWN PLATES; UNKNOWN RODS
Outcome Attributed to Adverse Event Required Intervention;
Patient Weight112
-
-