|
MEDOS INTERNATIONAL SÃ RL CH MOUNTAINEER OCT SPINAL SYSTEM MINI POLYAXIAL BONE SCREW 3.5 X 3.5 X 20MM; ORTHOSIS, CERVICAL PEDICLE SCREW SPINAL FIXATION
|
Back to Search Results |
|
| Model Number |
188318320 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Codes |
Failure of Implant (1924); Injury (2348)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Complainant part is not expected to be returned for manufacturer review/investigation.Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Event or Problem Description
|
|
It was reported that on (b)(6) 2020 the patient had a fall and fell face first resulting in the breaking of mountaineer screws at c1 and c2 midshaft.A revision surgery is pending the mri and ct results.Concomitant devices: unknown setscrews (part# unknown, lot# unknown, quantity unknown); unknown rods (part# unknown, lot# unknown, quantity unknown); unknown occipital plate (part# unknown, lot# unknown, quantity unknown).This report is for one (1) mountaineer oct spinal system mini polyaxial bone screw 3.5 x 3.5 x 20mm.This is report 2 of 4 for (b)(4).
|
| |
|
Additional Manufacturer Narrative
|
|
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Customer quality investigation:the complaint device was not received for investigation.The following investigation is based on the image(s) provided.The image(s) was reviewed, and the complaint condition could be confirmed as the x-ray image clearly displays that the screw was broken into 2+ pieces.Since the device was not returned, a dimensional inspection and a functional test were not able to be performed.A manufacturing record evaluation could not be performed as the lot number could not be determined from the image provided.A definitive assignable root cause could not be determined based on the provided information.During the investigation no product design issues or discrepancies were observed (based on the images) that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
| |
|
Search Alerts/Recalls
|
|
|