| Model Number |
MT22719-PNK |
| Medical Device Problem Code |
Wireless Communication Problem (3283)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/02/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that signal loss over one hour occurred.A review of the share logs was performed and signal loss was found within the investigation window.The allegation was confirmed.The probable cause was the transmitter and app were unable to establish a connection.No injury or medical intervention was reported.
|
| |
|