• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. SINGLE USE 3-LUMEN EXTRACTION BALLOON V

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. SINGLE USE 3-LUMEN EXTRACTION BALLOON V Back to Search Results
Model Number B-V243Q-A
Medical Device Problem Code Material Rupture (1546)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/23/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not returned to olympus for evaluation.The user¿s complaint was not confirmed.The user facility used one balloon that popped when trying to use it on a patient.A different balloon of the same model and lot was used and activated without it rupturing.The procedure was completed with no further issues.No findings available.The cause could not be determined.No further information was reported.A supplemental will be submitted if new information becomes available or the device is returned.
 
Event or Problem Description
The user facility reported that a balloon popped when trying to inflate it during a patient procedure.There was no patient injury or harm.No additional information was provided.This is file 2 of 3.
 
Additional Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The legal manufacturer reviewed the contents of this complaint.The legal manufacturer reported that since the actual product could not be confirmed and there were no abnormalities in the manufacturing records, the cause of the occurrence could not be identified.·in the manufacturing process of this product, we confirmed that all balloons were inflated, so it is believed that the balloons could be inflated without problems during shipping.From the results of past surveys, it is speculated that when the balloon was inflated and quarrying, the rubber of the balloon was damaged at the corner of the stone and the balloon ruptured.However, it was not possible to determine when and what load was applied and the balloon was punctured.In addition, the legal manufacturer reported that after checking the contents of the instruction manual (drawing number gk7054, revision number 8), the following caution was issued regarding the description related to this event.Confirm that the balloon is completely deflated when inserting the instrument into the endoscope.If the balloon is not deflated completely, the distal end of the instrument may extend from the distal end of the endoscope abruptly.This could cause patient injury, such as perforation, bleeding, or mucous membrane damage, and could damage the endoscope and/or instrument.Do not force to insert the instrument if resistance to insertion is encountered.Reduce the angulation (or lower the forceps elevator if applicable) until the instrument passes smoothly.The use of excessive force causes patient injury, such as perforation, bleeding, or mucous membrane damage.It may also damage the endoscope and/or instrument.Do not advance or extend the instrument abruptly.This could cause patient injury, such as perforation, bleeding, or mucous membrane damage.It could also damage the endoscope and/or instrument.Insert the instrument into the endoscope slowly.Otherwise, the instrument and/or the endoscope may be damaged.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE 3-LUMEN EXTRACTION BALLOON V
Common Device Name
SINGLE USE 3-LUMEN EXTRACTION BALLOON
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key10297688
Report Number8010047-2020-04618
Device Sequence Number811911
Product Code LQR
UDI-Device Identifier04953170368721
UDI-Public04953170368721
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
K091495
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 11/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberB-V243Q-A
Device Lot Number9ZV
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/23/2020
Supplement Date Received by Manufacturer10/09/2020
Initial Report FDA Received Date07/20/2020
Supplement Report FDA Received Date11/05/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-