| Model Number |
482802550 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Codes |
Injury (2348); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/25/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Device remains implanted in patient.
|
| |
|
Event or Problem Description
|
|
It was reported that an es2 integrated blade screw head fractured intra-operatively during a scheduled revision.While removing the screw on the bottom right side, the head of the screw broke off and the main body remained in the patient.The surgeon tried removing the rest of the screw but was unsuccessful.The surgeon had to close the wound with the broken screw left in the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Visual, dimensional, functional and material analysis could not be performed as the device was not returned because it remains implanted in the patient.A review of the device and complaint history records could not be performed as a valid lot number was not provided and could not be obtained.From es2 surgical technique: while the expected life of spinal implant components is difficult to estimate, it is finite.These components are made of foreign materials which are placed within the body for the potential fusion of the spine and reduction of pain.However, due to the many biological, mechanical and physicochemical factors which affect these devices but cannot be evaluated in vivo, the components cannot be expected to indefinitely withstand the activity level and loads of normal healthy bone.These implants are temporary internal fixation devices designed to stabilize the operative site during the normal healing process.After healing occurs, these devices serve no functional purpose and can be removed.Since the patient had fused, the implant had served its functional purpose.It is likely that the screw fractured due to fatigue on screw over the course of time.
|
| |
|
Event or Problem Description
|
|
It was reported that an es2 integrated blade screw head fractured intra-operatively during a scheduled revision.While removing the screw on the bottom right side, the head of the screw broke off and the main body remained in the patient.The surgeon tried removing the rest of the screw but was unsuccessful.The surgeon had to close the wound with the broken screw left in the patient.
|
| |
|