MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF Back to Search Results
Model Number 97715
Device Problems Overheating of Device (1437); Delayed Charge Time (2586); Charging Problem (2892)
Patient Problem Alteration In Body Temperature (2682)
Event Date 07/09/2020
Event Type  malfunction  
Manufacturer Narrative
Concomitant medical products: product id: 97755, serial#: (b)(4), product type: recharger.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a consumer regarding a patient who is implanted with a neurostimulator.It was reported that starting about a week prior to the report, confirmed as (b)(6) 2020, the patient has not been able to charge the implant and gets a poor recharge quality screen when trying to charge.The recharger paddle itself does not feel hot; however, the implant feels hot when charging.The patient noted that the sensation is under her skin at the implant site that gets hot, not the surface of her skin.The patient last noted having charged and felt stimulation about a week or maybe 5 days prior to the report.The patient was sent a replacement recharger to try and resolve the issue.There were no further complications reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTELLIS
Type of Device
STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
david gustafson
7000 central avenue ne rcw215
minneapolis, MN 55432
7635149628
MDR Report Key10299149
MDR Text Key199686567
Report Number3004209178-2020-12483
Device Sequence Number1
Product Code LGW
UDI-Device Identifier00643169781702
UDI-Public00643169781702
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 07/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date02/14/2019
Device Model Number97715
Device Catalogue Number97715
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/16/2020
Initial Date FDA Received07/20/2020
Date Device Manufactured02/26/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age71 YR
-
-