SMITH & NEPHEW, INC. TIBIAL SPACER BLOCK STANDARD; PRSTHSS, KNEE, PTLLFMRTBIAL, SM-CNSTRND, CMNTD, POLYMER/METAL/POLYMER
|
Back to Search Results |
|
| Model Number |
74012645 |
| Medical Device Problem Code |
Incomplete or Inadequate Connection (4037)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/25/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Event or Problem Description
|
|
It was reported that during surgery while using the block the spacer would not lock in.The surgeon said that he would simply hold it together for the minimal time needed and was able to move forward with the surgery.No back up item was needed, there was no delay to the case, there was no injury to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
The device used in treatment was returned for evaluation.Visual inspection of the returned device confirms that the device is broken and has severe signs of wear and tear from use.The device was manufactured in 2018.A review of the manufacturing records did not reveal any manufacturing or material abnormalities that could have caused or contributed to the reported incident.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.According to clinical/medical investigation, a review of this complaint case revealed there were no patient injuries reported.The procedure was completed using the same device.Since no patient injuries are being reported no further clinical/medical assessment is warranted at this time.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.The device is a reusable device that can be exposed to numerous surgeries and cleaning cycles.Probable causes that could contribute to the reported event, as reusable instruments are susceptible to damage from prolonged use and through misuse or rough handling.To avoid compromised performance or damage, it is recommended the device be inspected prior to and after each use and cleaning.We recommend that all re-usable instruments be routinely inspected for wear/damage and replaced as necessary.A review of risk management files and the instructions for use found that the reported failure was documented appropriately.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further actions are being taken at this time.We consider this investigation closed.Credit will be issued for the device.
|
| |
|
Search Alerts/Recalls
|
|
|