| Model Number |
MMT-551LNAP |
| Medical Device Problem Codes |
Connection Problem (2900); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/07/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the retainer ring was cracked and reservoir was not able to lock in place when inserted into pump.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received unable to perform the displacement test due to complete broken reservoir tube lip noted and also received with cracked battery tube threads, missing reservoir tube o-ring, cracked case reservoir tube window corner and missing end cap sticker.The p-cap not locks into the reservoir compartment properly due to completely broken reservoir tube lip noted.
|
| |
|