• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LP JUAREZ (MMJ) SHILEY; TUBE, TRACHEOSTOMY (W/WO CONNECTOR)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LP JUAREZ (MMJ) SHILEY; TUBE, TRACHEOSTOMY (W/WO CONNECTOR) Back to Search Results
Model Number 8PERC
Medical Device Problem Code Use of Device Problem (1670)
Health Effect - Clinical Code Extubate (2402)
Date of Event 07/01/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, the trach hardware was broken and trach plate identified to be sutured and remained in place.Cardiopulmonary resuscitation(cpr) initiated with return of spontaneous circulation(rosc) within 2 minutes.Patient was subsequently intubated with endotracheal tube(ett), defective trach removed, packaging sequestered and new percutaneous trach reinserted.In addition the patient had pre-existing medical conditions like hypertension and prior stroke with mild residual weakness admitted with acute left middle cerebral artery syndrome(mca) stroke in setting of left supraclinoid internal carotid artery(ica) occlusion.
 
Additional Manufacturer Narrative
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was available for evaluation.Visual inspection noted the head on the product was disconnected.It was reported that the flange was broken.The reported issue was confirmed.The most likely cause was traced to a component failure.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.A secondary review of the device history records found no potentially contributing factors.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
According to the reporter, post-insertion of the tracheostomy tube, the device's hardware was found broken.The healthcare professional noted that the patient was desaturated concurrent with inner cannula and the external hardware of the device actively advancing outward.The trach plate identified to be sutured and remained in place.Cpr was initiated with rosc within 2 minutes.Patient had a new percutaneous trach reinserted.
 
Additional Manufacturer Narrative
The incident device has been received and is under evaluation.When the device evaluation is complete a follow-up report will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHILEY
Common Device Name
TUBE, TRACHEOSTOMY (W/WO CONNECTOR)
Manufacturer (Section D)
COVIDIEN LP JUAREZ (MMJ)
9560 jose rodriguez drive
el paso TX 79927
MDR Report Key10302615
Report Number2936999-2020-00510
Device Sequence Number2576843
Product Code BTO
UDI-Device Identifier30884522006709
UDI-Public30884522006709
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
K884730
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/23/2024
Device Model Number8PERC
Device Catalogue Number8PERC
Device Lot Number19D0644JZX
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/05/2020
Initial Date Received by Manufacturer 07/02/2020
Supplement Date Received by Manufacturer08/17/2020
09/03/2020
Initial Report FDA Received Date07/21/2020
Supplement Report FDA Received Date09/02/2020
09/22/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age52 YR
Patient Weight99
-
-