| Model Number |
2800-45 |
| Medical Device Problem Code |
Use of Device Problem (1670)
|
| Health Effect - Clinical Code |
Needle Stick/Puncture (2462)
|
| Date of Event |
06/28/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
As part of the investigation, instrumentation laboratory (il) reviewed the online help documentation (acl top family 50 series operator's manual).The investigation confirmed that there are appropriate instructions and cautions regarding the activities of cleaning and removing cuvettes from the waste area of the instrument.Per the acl top family 50 series operator's manual, the operator action to address the two alarms presented during the incident is to perform a recovery and if the alarm persists to call for service.Il confirmed that the laboratory technician did not call il service or the technical service helpline while performing these activities.The acl top 550 cts performed as intended with no malfunction and its labeling provides appropriate instructions and cautions to the user.Therefore, no remedial action is indicated.
|
| |
|
Event or Problem Description
|
|
It was reported that an acl top 550 cts instrument operator developed a laceration resulting in bleeding on the finger of her gloved hand.This incident occurred while troubleshooting to clean and remove cuvettes from the waste area of the instrument.The staff member received medical attention in the emergency department at the hospital and filed an incident report.No stitches were necessary.The staff member was placed on medication for (b)(6) prophylaxis.
|
| |
|