Model Number J546G
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/22/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.Attempts are being made to receive a device for evaluation.To date the device has not been returned.If the device or further details are received at the later date a supplemental medwatch will be sent.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Events were submitted via 2210968-2020-05439, 2210968-2020-05440, 2210968-2020-05441 and 2210968-2020-05442.
 
Event or Problem Description
It was reported that the patient underwent an unknown surgery on (b)(6) 2020 and the suture was used.During the surgery, suture was broken in a row during continuous suturing.Another like suture with another lot number was used with no problem.There were no adverse consequences to the patient.
 
Additional Manufacturer Narrative
(b)(4).Date sent to the fda: 09/15/2020.Additional h3 investigation summary:: it was reported performance breakage suture.An opened box and with two unopened samples of product code j546g, lot # pmz360 were returned for analysis.During the visual inspection of two unopened samples, no defects were found on the packages.The samples were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand and no defects, damaged or suture breakage were noted.A functional test was performed and the tensile strength force was above the minimum requirement.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Per the condition of samples received, no performance - breakage suture was found and the tested samples met the finished goods requirements.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VCL CT BRD VIO 18IN 7-0 D/A TG140-8
Common Device Name
SUTURE, ABSORBABLE, SYNTHETIC
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876
MDR Report Key10305507
Report Number2210968-2020-05441
Device Sequence Number11479312
Product Code GAM
UDI-Device Identifier10705031041097
UDI-Public10705031041097
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K022269
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2024
Device Model NumberJ546G
Device Catalogue NumberJ546G
Device Lot NumberPMZ360
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/24/2020
Initial Date Received by Manufacturer 07/02/2020
Supplement Date Received by Manufacturer08/24/2020
Initial Report FDA Received Date07/21/2020
Supplement Report FDA Received Date09/15/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-