|
(b)(4).A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.Attempts are being made to receive a device for evaluation.To date the device has not been returned.If the device or further details are received at the later date a supplemental medwatch will be sent.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Events were submitted via 2210968-2020-05439, 2210968-2020-05440, 2210968-2020-05441 and 2210968-2020-05442.
|
|
(b)(4).Date sent to the fda: 09/15/2020.Additional h3 investigation summary: it was reported performance breakage suture.An opened box and with two unopened samples of product code j546g, lot # pmz360 were returned for analysis.During the visual inspection of two unopened samples, no defects were found on the packages.The samples were opened, and the swage and attachment area were noted to be as expected.The sutures were dispensed without problems and examined along of the strand and no defects, damaged or suture breakage were noted.A functional test was performed and the tensile strength force was above the minimum requirement.The manufacturing records were reviewed, and the manufacturing/packaging criteria were met prior to the release of this batch.Per the condition of samples received, no performance - breakage suture was found and the tested samples met the finished goods requirements.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon inc, or its employees that the report constitutes an admission that the product, ethicon inc or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|