| Model Number |
N051 |
| Medical Device Problem Code |
Premature Discharge of Battery (1057)
|
| Health Effect - Clinical Code |
No Code Available (3191)
|
| Date of Event |
11/30/2019
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Product is not expected to be returned.As no further information concerning this report is expected, our investigation is complete.This investigation will be updated should further pertinent information be provided.(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that this device recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.This device was explanted and is not expected to be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
The returned implantable pulse generator was analyzed and a review of the device memory confirmed that a low voltage alert, code 1003, was recorded.The battery voltage was lower than expected, but still supported full device function.Using historical daily battery voltage measurement data, engineers determined that this device was demonstrating behavior consistent with a high current condition associated with a compromised low voltage capacitor connected to the device's battery.Low voltage capacitors are used in the device's high voltage charging operation in order to facilitate fast charge times.The behavior of these capacitors resulted in a high current drain, which was depleting this device's battery faster than normal.Boston scientific has issued an advisory communication regarding an older subset of cognis/teligen devices that is more susceptible to this anomaly.Specifically, the performance of a low voltage capacitor may be compromised over time, causing an increased current drain that can lead to premature battery depletion.This particular device was not included in the advisory population.Patient code 3191 captures the reportable event of surgery.
|
| |
|
Event or Problem Description
|
|
It was reported that this device recorded a code 1003 indicative of battery voltage too low for the projected remaining capacity.Data analysis was not performed for this device prior to explant procedure.The device was explanted and replaced due to reported premature battery depletion and code 1003.This device was returned for analysis.The product was analyzed and the premature discharge of battery and message - error code 1003 allegations were confirmed.No other suspicious faults or resets were noted.Based on the memory log the device battery voltage and the power usage was normal while implanted, all therapy was available while implanted.
|
| |
|