|
Device passed the functional test, including the displacement test, rewind, basic occlusion test, occlusion test, prime/a33 test, excessive no delivery test and delivery accuracy test at 0.08740 inches.No over delivery anomaly or under delivery anomaly noted.The stop current and run current measurement tests are within specification.Device also passed self test, off no power alarm test and a21 error test.Device had intermittent button response on the esc button due to flattened dome switch.No button error alarm noted.During visual inspection, the keypad connector on the lcd board was found locked properly.Device uploaded properly using care link.Device had cracked battery tube threads, scratched reservoir tube window and cracked reservoir tube lip.The test p-cap and reservoir does lock in place in the reservoir compartment.The over delivery anomaly was not specified in the customer notes.Unable to verify over delivery anomaly.However, from analyzing the daily insulin total usage a week prior to the event date, there was no anomaly noted.The daily insulin totals were between 66 unit to 73 units with the exception on (b)(6) 2020 was only basal delivery.There was no obvious increase of insulin used.
|