| Model Number |
MMT-551NAB |
| Medical Device Problem Code |
Priming Problem (4040)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/15/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump was squirting out insulin very fast during priming and was not like normal.No harm requiring medical intervention was reported.The insulin pump will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed the displacement test but prime or fill anomaly during the basic occlusion test due to loose drive support disk.Unable to perform the occlusion, prime/a33 and excessive no delivery test due to loose drive support disk.(b)(4).
|
| |
|