BARD ACCESS SYSTEMS HEMOSPLIT LONG-TERM HEMODIALYSIS CATHETER, POLYURETHANE, KIT, 14.5F
|
Back to Search Results |
|
| Model Number |
5703690 |
| Medical Device Problem Code |
Fracture (1260)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/24/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
As the lot number for the device was provided, a manufacturing review will be performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.Expiry date (02/2022).
|
| |
|
Event or Problem Description
|
|
It was reported that some post catheter implantation in right jugular vein, the needle joint allegedly crack.There was no reported patient injury.
|
| |
|
Event or Problem Description
|
|
It was reported that during catheter implantation in right jugular vein, the needle joint allegedly cracked.The procedure was completed using another device.There was no reported patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: manufacturing review: a lot history review was performed.This is the only complaint to date for this lot number.Therefore, a device history record review is not required.Investigation summary: a sample evaluation could not be performed, as the device was not returned.Therefore, the report of needle joint crack cannot be confirmed.A definitive root cause could not be determined based upon available information.Labeling review: a review of product labeling documentation (e.G.Procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4(expiry date :02/2022),g4.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
| |
|