| Model Number |
MMT-723LNAP |
| Medical Device Problem Codes |
Device Difficult to Program or Calibrate (1496); Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
05/15/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump's buttons were harder to press.Customer stated the insulin pump had small crack on the corner of screen and rainbow effect on screen.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed displacement test and self test.Device received with all buttons responding properly.No damage on keypad assembly.No unlocked lcd keypad connector.Device received with cracked lcd window, minor scratched lcd window, cracked case from display window corner, cracked case and cracked reservoir tube lip.The test infusion set and reservoir does lock into place.(b)(4).
|
| |
|