• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. EMPOWR PS KNEE; EMPOWR PS KNEETM TIBIAL INSERT, SIZE 5 BRIDGE UP, 10MM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. EMPOWR PS KNEE; EMPOWR PS KNEETM TIBIAL INSERT, SIZE 5 BRIDGE UP, 10MM Back to Search Results
Model Number 343-10-755
Health Effect - Clinical Code Fall (1848)
Date of Event 06/24/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reason for this revision surgery was reported as distal femur fracture after a fall.The previous surgery and the surgery detailed in this event occurred 16 days apart.The healthcare professional indicated there was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the device history record (dhr) show that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There were no non-conforming material reports (ncmr) associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are needing a review.The root cause of this complaint was a revision surgery due to distal femur fracture after a fall.There were no findings during this evaluation that indicate the reported device was defective.No other information was submitted with the complaint regarding pre-existing conditions of the patient or any activities the patient was involved in that may have contributed to the event.There are multiple factors that may contribute to an event that are outside the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Event or Problem Description
Revision surgery - patient fell and caused a distal femur fracture, a femoral placed and new insert was placed.
 
Additional Manufacturer Narrative
Additional narrative: patient fell and caused a distal femur fracture.A femoral rod was placed and new insert was placed.Additional reporting on this event will be provided as a supplemental report to this document if it becomes available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EMPOWR PS KNEE
Common Device Name
EMPOWR PS KNEETM TIBIAL INSERT, SIZE 5 BRIDGE UP, 10MM
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin, tx 78758-5445,
MDR Report Key10307539
Report Number1644408-2020-00576
Device Sequence Number13659335
Product Code JWH
UDI-Device Identifier00888912171960
UDI-Public(01)00888912171960
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K160342
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date02/28/2024
Device Model Number343-10-755
Device Catalogue Number343-10-755
Device Lot Number249U1036
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/24/2020
Supplement Date Received by Manufacturer06/24/2020
Initial Report FDA Received Date07/22/2020
Supplement Report FDA Received Date10/06/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age76 YR
Patient Weight71
-
-