MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE BAL SIZER DISTRACTOR; ATTUNE INSTRUMENTS : SIZING/MEASURING INSTRUMENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE BAL SIZER DISTRACTOR; ATTUNE INSTRUMENTS : SIZING/MEASURING INSTRUMENTS Back to Search Results
Model Number 2544-00-519
Medical Device Problem Codes Device-Device Incompatibility (2919); Naturally Worn (2988)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/06/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Prior to the case, the scrub tech noticed the balanced sizing tool from a set of loaner instruments was not working properly.No delays.
 
Event or Problem Description
Additional information received states that instruments look like either due to wear and tear or the piece itself wasn¿t machined properly.It¿s difficult to get the two-part to seat properly.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : examination of the returned device revealed the mating components of this part are functional, however, significant resistance during operation is noted.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE BAL SIZER DISTRACTOR
Common Device Name
ATTUNE INSTRUMENTS : SIZING/MEASURING INSTRUMENTS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key10307799
Report Number1818910-2020-16493
Device Sequence Number16948260
Product Code HWT
UDI-Device Identifier10603295384403
UDI-Public10603295384403
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2544-00-519
Device Catalogue Number254400519
Device Lot NumberABA91778
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/27/2020
Initial Date Received by Manufacturer 07/06/2020
Supplement Date Received by Manufacturer07/27/2020
07/30/2020
Initial Report FDA Received Date07/22/2020
Supplement Report FDA Received Date07/29/2020
07/31/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-