STRYKER ORTHOPAEDICS-MAHWAH DETACHABLE FLEX SHAFT; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 2107-2200
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/03/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
During total hip replacement surgery, the flex.Drill guide was broken while drilling for acetabular cup screws.20 min delay, right side."during the mounting process of total hip prothesis while screwing acetabular cup, flex drill guide is broken".
 
Additional Manufacturer Narrative
Reported event: an event regarding crack/fracture involving a trident driver shaft was reported.The event was confirmed by analysis of the returned device.Method & results: device evaluation and results: visual inspection: visual inspection was performed as part of the material evaluation and indicated the following comments: the shaft was examined with the aid of a stereo-microscope at magnifications up to 50x.The distal fracture surface was examined using a scanning electron microscope (sem).Dimensional inspection: not performed as the reported device was returned damaged and in its current condition would not be an accurate reflection of its original manufactured condition.Functional inspection: not performed as the reported device was returned damaged and in its current condition would not be an accurate reflection of its original manufactured condition.Material analysis: material evaluation was completed and indicated the following comments: the shaft was examined with the aid of a stereo-microscope at magnifications up to 50x.The distal fracture surface was examined using a scanning electron microscope (sem).The fracture surface morphology was consistent with torsional overload.Energy dispersive spectrscopy showed that the base material of the flex shaft was consistent with 17-7 stainless steel.Based on the given information no materials discrepancies were observed on the surfaces examined.Clinician review: no medical records were received for review with a clinical consultant.Device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there has been no other similar event for the reported lot.Conclusion: the investigation concluded, as a result of analysis of the device "the fracture surface morphology was consistent with torsional overload.Energy dispersive spectroscopy showed that the base material of the flex shaft was consistent with 17-7 stainless steel.Based on the given information no materials discrepancies were observed on the surfaces examined.No further investigation is required at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
 
Event or Problem Description
During total hip replacement surgery, the flex.Drill guide was broken while drilling for acetabular cup screws.20 min delay, right side "during the mounting process of total hip prothesis while screwing acetabular cup, flex drill guide is broken".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DETACHABLE FLEX SHAFT
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key10307928
Report Number0002249697-2020-01525
Device Sequence Number3371277
Product Code LPH
UDI-Device Identifier07613327212327
UDI-Public07613327212327
Combination Product (Y/N)N
PMA/510(K) Number
K161569
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2023
Device Model Number2107-2200
Device Catalogue Number2107-2200
Device Lot NumberBDCAP
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/23/2020
Initial Date Received by Manufacturer 07/04/2020
Supplement Date Received by Manufacturer09/23/2020
Initial Report FDA Received Date07/22/2020
Supplement Report FDA Received Date10/20/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age54 YR
Patient Weight75
-
-