• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER ORTHOPAEDICS-MAHWAH TRIDENTII TRITANIUM CLUSTER48D; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ORTHOPAEDICS-MAHWAH TRIDENTII TRITANIUM CLUSTER48D; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED Back to Search Results
Model Number 702-04-48D
Medical Device Problem Codes Insufficient Device Problem Information (3190); Difficult or Delayed Separation (4044)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/07/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.It was noted that the device is not available for evaluation.Should additional information become available, it will be provided in a supplemental report upon completion of the investigation.
 
Event or Problem Description
Primary procedure, right hip.It was reported that the impactor bolt became stuck in the shell, and that the impactor sheared off where the handle meets the curved part of the impactor.All devices (including the shell) were removed from the patient and another shell was implanted.Surgery was completed successfully with a delay of approximately 2 minutes.Rep may be able to provide some additional patient information, otherwise confirmed that no further information will be released by the hospital or surgeon.
 
Additional Manufacturer Narrative
Reported event: an event regarding a disassembly issue involving a trident shell and an impactor bolt was reported.The event was not confirmed.Method & results: device evaluation and results: not performed as product was not returned.Clinician review: no medical records were received for review with a clinical consultant.Device history review: all devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the reported lot.Conclusion: the event itself could not be confirmed and exact cause of the event could not be determined because insufficient information was provided.Further information such as product return as well as images of the device taken when the alleged event was identified is required to complete the investigation for confirming the event and determining a root cause.If further information becomes available or the product is returned, this investigation will be re-opened.H3 other text : device not returned.
 
Event or Problem Description
Primary procedure, right hip.It was reported that the impactor bolt became stuck in the shell, and that the impactor sheared off where the handle meets the curved part of the impactor.All devices (including the shell) were removed from the patient and another shell was implanted.Surgery was completed successfully with a delay of approximately 2 minutes.Rep may be able to provide some additional patient information, otherwise confirmed that no further information will be released by the hospital or surgeon.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRIDENTII TRITANIUM CLUSTER48D
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED
Manufacturer (Section D)
STRYKER ORTHOPAEDICS-MAHWAH
325 corporate drive
mahwah NJ 07430
MDR Report Key10307934
Report Number0002249697-2020-01526
Device Sequence Number13667910
Product Code LPH
UDI-Device Identifier07613327343861
UDI-Public07613327343861
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
PMA/510(K) Number
K171768
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,other
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 01/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date01/31/2025
Device Model Number702-04-48D
Device Catalogue Number702-04-48D
Device Lot Number76857301A
Was Device Available for Evaluation? No
Date Returned to Manufacturer07/13/2020
Initial Date Received by Manufacturer 07/07/2020
Supplement Date Received by Manufacturer01/05/2021
Initial Report FDA Received Date07/22/2020
Supplement Report FDA Received Date01/13/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other; Required Intervention;
Patient Age76 YR
-
-