Model Number 1012453-15 |
Device Problem
Material Separation (1562)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 07/03/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device is expected to be returned for investigation.It has not yet been received.A follow-up report will be submitted with all additional relevant information.
|
|
Event Description
|
It was reported that after removing the balloon protective sheath without resistance, the 4.0x15mm nc trek balloon was separated.There was no patient involvement.A new 4.0x15mm nc trek balloon was used to successfully continue the procedure.There were no adverse patient effects and no clinically significant delay in the procedure.No additional information was provided.
|
|
Manufacturer Narrative
|
The device was returned for analysis.The reported material separation was able to be confirmed.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history identified no other similar incidents from this lot.The investigation determined the reported difficulties appear to be related to circumstances of the procedure.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
|