• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. SENSOR ENLITE MMT-7008A; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. SENSOR ENLITE MMT-7008A; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-7008A
Medical Device Problem Codes Incorrect, Inadequate or Imprecise Result or Readings (1535); Unexpected Therapeutic Results (1631)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/15/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
Information received by medtronic the user reported inaccurate sensor readings that triggered a threshold suspend.The customer¿s blood glucose was 70 mg/dl and sensor glucose was 40 mg/dl at the time of the event, the difference is outside the acceptable range.Suspend on low limit in sensor settings was at 70 mg/dl.No harm requiring medical intervention was reported.The sensor will be returned for analysis.
 
Additional Manufacturer Narrative
Inspected one opened and used sensor and performed continuity resistance test and sensor failed per specifications.Found sensor electrode out of cannula tubing.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSOR ENLITE MMT-7008A
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key10308061
Report Number2032227-2020-150860
Device Sequence Number10968882
Product Code OZO
UDI-Device Identifier20643169786725
UDI-Public(01)20643169786725(17)201130(10)HG4C7YE
Combination Product (Y/N)N
Initial Reporter StateHI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Expiration Date11/30/2020
Device Model NumberMMT-7008A
Device Catalogue NumberMMT-7008A
Device Lot NumberHG4C7YE
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/10/2020
Initial Date Received by Manufacturer 07/16/2020
Supplement Date Received by Manufacturer08/14/2020
Initial Report FDA Received Date07/22/2020
Supplement Report FDA Received Date08/16/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age77 YR
Patient Weight61
-
-