| Model Number |
1550350-23 |
| Medical Device Problem Code |
Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
06/04/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device location is unknown as account could not be determined/reached.Investigation is not yet complete.A follow-up report will be submitted with all relevant information.
|
| |
|
Event or Problem Description
|
|
User facility medwatch report received that states stent was not on balloon and properly secured to the stent delivery system.Rt caught the error prior to using on patient.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot specific quality issue.The investigation was unable to determine a conclusive cause for the reported stent dislodgement.There is no indication of a product quality issue with respect to manufacture, design or labeling.
|
| |
|