• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE PS FEM TRIAL SZ 4 RT; ATTUNE INSTRUMENTS : FEMORAL TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE PS FEM TRIAL SZ 4 RT; ATTUNE INSTRUMENTS : FEMORAL TRIALS Back to Search Results
Model Number 2545-00-724
Medical Device Problem Code Break (1069)
Health Effect - Clinical Codes Bone Fracture(s) (1870); Not Applicable (3189)
Date of Event 01/13/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: examination of the returned device confirmed the reported event.Examination of the returned device confirmed the trial had broken into two pieces.The trial is fractured across the medial condyle.All pieces were returned for evaluation.The device exhibited minimal wear depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Event or Problem Description
During impaction the lateral condyle of the attune femoral trial broke off (254500724).As the pieces of the broken trial were removed it was found that the patient had developed a non displaced lateral femoral condylar fracture starting at the anterior portion of the box resection and extending to the posterior portion of the femur.This was repaired using two 4.0mm cannulated screws and the surgery was finished with no further incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE PS FEM TRIAL SZ 4 RT
Common Device Name
ATTUNE INSTRUMENTS : FEMORAL TRIALS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key10309167
Report Number1818910-2020-16542
Device Sequence Number3246828
Product Code HWT
UDI-Device Identifier10603295133964
UDI-Public10603295133964
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 07/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2545-00-724
Device Catalogue Number254500724
Device Lot NumberMVMBWY980
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/16/2020
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 07/13/2020
Initial Report FDA Received Date07/22/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age62 YR
-
-