| Model Number |
ZCT400 |
| Medical Device Problem Code |
Unintended Movement (3026)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); Halo (2227); No Code Available (3191)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown, not provided, but the best estimate date is between (b)(6) 2020.(b)(4).All pertinent information available to johnson and johnson surgical vision has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that surgeon has a patient with unexpected refractive outcome in the left eye, status post uncomplicated phaco with a toric posterior chamber intraocular lens placement, and subsequent exchange with another toric lens.The patient had irregular astigmatism.Initial surgery was performed on (b)(6) 2020 and lens was noted to be in alignment during immediate post-operative course.Post-op day 30, the patient was found to have toric misalignment and surgeon¿s repositioning revealed that a higher power toric was indicated.Patient agreed to lens replacement and the lens exchange was performed without complication in secondary surgical procedure on (b)(6) 2020.Resultant uncorrected vision was poor, and refraction reveals high post-operative cylinder requirement.No corneal or macular edema or evidence of neuropathy or inflammation.Additional information was received and it was learnt that patient was also experiencing halos, blurry vision, and glare after the initial surgery.There was also age related nuclear cataract observed.There was tecnis toric rotation of more than 10 degrees.Patient was also prescribed eye drops.No further information provided.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: the product testing could not be performed as the product was not returned.The reported complaint cannot be confirmed.Manufacturing records review: the manufacturing records for the device were reviewed.The product was manufactured and released according to specification.A search in complaint system revealed no other complaints has been received for this production order number.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|