• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OGDEN MANUFACTURING PLANT OPTIFLUX F18NRE DIALYZER FINISHED ASSY; DIALYZER, CAPILLARY, HOLLOW FIBER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OGDEN MANUFACTURING PLANT OPTIFLUX F18NRE DIALYZER FINISHED ASSY; DIALYZER, CAPILLARY, HOLLOW FIBER Back to Search Results
Model Number 0500308E
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 07/06/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The plant investigation is in process.A supplemental mdr will be submitted upon completion of this activity. .
 
Event or Problem Description
A user facility nurse reported that a dialyzer blood leak occurred 5 minutes after the initiation of the patient¿s hemodialysis (hd) treatment.The blood leak was noted as being an internal blood leak.The machine, a fresenius 4008s machine, alarmed appropriately with a blood leak alarm.Blood leak test strips were not used as the leak was observed in the arterial head cap of the dialyzer.There was no defect or damage noted on the dialyzer.The patient¿s estimated blood loss (ebl) was approximately 250 ml.There was no patient injury, adverse events, or medical intervention required as a result of this event.The patient was restarted on the same machine and treatment completed successfully with new supplies.The complaint device was is not available to be returned to the manufacturer for evaluation because it was discarded.
 
Additional Manufacturer Narrative
Plant investigation: the reported complaint was not confirmed as the complaint device was not returned for manufacturer evaluation.A photo of a dialyzer connected to a machine was provided.The dialyzer in the photo had blood present within the dialyzer fibers (as expected).A small amount of blood is visible outside of the fibers in the bell area of the housing.A production records review was performed on the reported lot.An investigation of the device history records (dhr) was conducted by the manufacturer.There was no indication of product nonacceptance, deviation, non-conformance, rework, labeling or process control failure during the manufacturing process which could be associated with the reported event.The lot met all release criteria.Although a photo of the dialyzer was provided, a definitive conclusion regarding the complaint incident cannot be reached without physical examination of the actual device.Therefore, the complaint is not confirmed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OPTIFLUX F18NRE DIALYZER FINISHED ASSY
Common Device Name
DIALYZER, CAPILLARY, HOLLOW FIBER
Manufacturer (Section D)
OGDEN MANUFACTURING PLANT
director, site quality
475 west 13th street
ogden UT 84404
MDR Report Key10310630
Report Number1713747-2020-00297
Device Sequence Number13667909
Product Code FJI
UDI-Device Identifier00840861100088
UDI-Public00840861100088
Combination Product (Y/N)N
Initial Reporter CountryMX
PMA/510(K) Number
K043244
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 07/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/30/2021
Device Model Number0500308E
Device Catalogue Number0500308E
Device Lot Number18LU01001
Was Device Available for Evaluation? No
Device AgeMO
Initial Date Received by Manufacturer 07/06/2020
Supplement Date Received by Manufacturer07/28/2020
Initial Report FDA Received Date07/22/2020
Supplement Report FDA Received Date07/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FRESENIUS 4008S MACHINE
Patient Age32 YR
Patient Weight98
-
-