According to the reporter, the catheter had a leak from the "wavy piece" when used at high pressure while the machine and blood pump was running.It was mentioned that the catheter had sucked in air and blood had leaked from the back.The catheter was repaired, they were using a 23 but then used a 28 kit to repair it.There was no reported patient outcome.
|
|
Evaluation summary: medtronic conducted an investigation based upon all information received.The device was not returned, but a photo was available for evaluation.Visual inspection noted the first image depicts the catheter inserted into a patient and there is blood at the point of insertion as well as on the catheter itself.The second image depicts the hub, extension tubes and clamps, and there is blood on the catheter and the extension tubes are both filled with blood.The third image depicts a finger pointing a hand written piece of paper.It was reported that the bifurcate was cracked, broken or leaking.The reported issue could not be confirmed.The most likely cause could not be established from the information available.The manufacturing records for each device are thoroughly reviewed prior to release to ensure that it meets all medtronic quality specifications.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|