| Model Number |
6215 |
| Medical Device Problem Codes |
Break (1069); Burst Container or Vessel (1074); Device Contamination with Body Fluid (2317); Activation, Positioning or Separation Problem (2906)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
01/16/2018
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Product event summary: the balloon catheter was returned and analyzed.The analysis indicated that the mechanical operation of the balloon catheter was occluded.There was an issue with the catheter shaft.The balloon was ruptured.Blood was observed on the shaft of the delivery catheter.Visual analysis of the balloon catheter indicated damage during use.The analyst noted the balloon catheter was returned without the inflation syringe and the coronary wire.There is blood at the distal tip orifice of the balloon catheter.The balloon is ruptured on the balloon catheter.A guidewire was inserted into the balloon catheter and came to an occlusion at 39 cm.Blood and contrast occluded the shaft of the balloon catheter.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that during an implant procedure, the balloon catheter would not allow the coronary wire to pass through and exit the catheter lumen.The catheter was not used and another was used.The catheter was returned to the manufacturer, analyzed, and tested out of specification.No patient complications have been reported as a result of this event.
|
| |
|