ST. JUDE MEDICAL PACEL BIPOLAR PACING CATHETER; ELECTRODE, PACEMAKER, TEMPORARY Back to Search Results
Model Number 401771
Medical Device Problem Code Material Separation (1562)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/15/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event or Problem Description
Four days following the placement of the temporary electrode, the proximal leg wire of the device broke requiring the device to be replaced.The patient's treatment was completed with the replacement device and the patient was released.
 
Additional Manufacturer Narrative
Because the reported catheter was not returned, investigation was limited to visual inspection of a photograph received from the event report.The leg wire appeared to be exposed and detached from the catheter assembly.Visual inspection was based solely upon a review of the photographs provided.The device was not returned.Additional investigation revealed the braided wire separated from the leg wire tube and the braided wire appears to be fractured close to the bifurcation cap.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported event remains unknown.
 
Event or Problem Description
One day after the placement of the temporary electrode, the proximal leg wire of the device broke requiring the device to be replaced.The patient's treatment was completed with the replacement device and the patient was released.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PACEL BIPOLAR PACING CATHETER
Common Device Name
ELECTRODE, PACEMAKER, TEMPORARY
Manufacturer (Section D)
ST. JUDE MEDICAL
14901 deveau place
minnetonka MN 55345 2126
MDR Report Key10311062
Report Number2182269-2020-00064
Device Sequence Number824881
Product Code LDF
UDI-Device Identifier05414734001137
UDI-Public05414734001137
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K152784
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2022
Device Model Number401771
Device Catalogue Number401771
Device Lot Number7269362
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 06/22/2020
Supplement Date Received by Manufacturer07/24/2020
Initial Report FDA Received Date07/22/2020
Supplement Report FDA Received Date07/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-