• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 11MM/125 DEG TI CANN TFNA 360MM/RIGHT - STERILE; ROD,FIXATION,INTRAMEDULLARY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 11MM/125 DEG TI CANN TFNA 360MM/RIGHT - STERILE; ROD,FIXATION,INTRAMEDULLARY Back to Search Results
Model Number 04.037.126S
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 01/01/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Event year is reported as 2020; however exact date of event is unknown.Reporter is jnj representative.The investigation could not be completed, no product was received; no conclusion could be drawn at the time of filing this report.The device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on an unknown date, patient underwent removal of proximal femoral nailing system (tfna) nail that was broken after 6 months of implantation.Originally implanted with tfna nail, tfna blade and unknown locking screws on an unknown date.The fenestrated helical blade was intact, there was indentations on the blade due to the nail breaking around it.Fragments were completely removed.Removal was done fairly easily without significant intervention, then a total hip arthroplasty was completed.There was patient consequence reported.Concomitant device reported: unknown locking screw (part# unknown, lot# unknown, quantity 1).This complaint involves 2 devices.This report is for (1) 11mm/125 deg ti cann tfna 360mm/right - sterile.This is report 1 of 2 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h4, h6- the 1mm/125 deg ti cann tfna 360mm/right - sterile (p/n: 04.037.126, lot number: unk) was received at us cq.Visual inspection of the complaint device showed the nail had broken at the proximal slot.Drill marks were observed inside the slot.This complaint is confirmed as the nail has broken at the proximal slot and shows drill marks.Pie has been launched to address the identified issue.The need for further corrective/preventive action will be assessed within the pie.Further investigation will not be conducted in this complaint as it will be addressed within pie no definitive root cause could be determined based on the provided information.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
11MM/125 DEG TI CANN TFNA 360MM/RIGHT - STERILE
Common Device Name
ROD,FIXATION,INTRAMEDULLARY
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key10311139
Report Number2939274-2020-03276
Device Sequence Number11340362
Product Code HSB
UDI-Device Identifier10886982096401
UDI-Public(01)10886982096401
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K160167
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number04.037.126S
Device Catalogue Number04.037.126S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/16/2020
Initial Date Received by Manufacturer 07/03/2020
Supplement Date Received by Manufacturer08/05/2020
Initial Report FDA Received Date07/22/2020
Supplement Report FDA Received Date08/20/2020
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
TFNA FENESTRATED HELICAL BLADE 105MM - STERILE; UNK - SCREWS: LOCKING
-
-