Model Number 657310S
Medical Device Problem Codes Positioning Failure (1158); Material Fragmentation (1261)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/29/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Device will not be returned.If additional information becomes available, it will be provided in a supplemental report.Device discarded.
 
Event or Problem Description
It was reported: when the locking screw was inserted into the plate, the screw thread was scraped and it could not be locked.Collect the screw and debris.The surgeon inserted a new screw and finished the operation successfully.
 
Event or Problem Description
It was reported: when the locking screw was inserted into the plate, the screw thread was scraped and it could not be locked.Collect the screw and debris.The surgeon inserted a new screw and finished the operation successfully.
 
Additional Manufacturer Narrative
The reported event could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.A device inspection was not possible since the affected device was not returned and no other evidences were provided for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.More information as well as the affected device must be available in order to determine the exact root cause of the failure.However, based on past complaint history some of the probable causes could be, but not limited to: overtightening of the screw, insertion of the screw in an inadequate angle etc.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LOCKING SCREW VARIAX2 T10, FULL THREAD, 3.5MM / L10MM
Common Device Name
PLATE, FIXATION, BONE
Manufacturer (Section D)
STRYKER GMBH
bohnackerweg 1
postfach
selzach 2545
SZ  2545
MDR Report Key10311168
Report Number0008031020-2020-01946
Device Sequence Number860524
Product Code HRS
UDI-Device Identifier07613327086799
UDI-Public07613327086799
Combination Product (Y/N)N
PMA/510(K) Number
K132502
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 10/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/30/2023
Device Model Number657310S
Device Catalogue Number657310S
Device Lot NumberY54933
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/29/2020
Supplement Date Received by Manufacturer09/24/2020
Initial Report FDA Received Date07/22/2020
Supplement Report FDA Received Date10/20/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age53 YR
-
-