• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. DRILL TIP GUIDE WIRE 2.4MM; PASSER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. DRILL TIP GUIDE WIRE 2.4MM; PASSER Back to Search Results
Model Number 014396
Medical Device Problem Code Difficult to Open or Remove Packaging Material (2922)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/02/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during procedure, when the package was opened and the rubber protective cap was removed from the rod, a piece of the cap was completely adhered to the rod.A backup device was available to complete the procedure with no delay or patient injuries.
 
Additional Manufacturer Narrative
The reported 2.4 mm drill tipped guide wire, intended for use in treatment, will not be returned for evaluation.Without the reported product a visual evaluation cannot be performed and the customers complaint cannot be confirmed, however, a trend of this nature has been observed with this product in the field, prompting a design change which has been implemented to remove the silicone cap and replace it with a tpu sleeve to prevent this issue going forward.A review of the manufacturing and complaint records was performed for the reported lot, there were no indications that would suggest that the device did not meet product specifications upon release into distribution.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRILL TIP GUIDE WIRE 2.4MM
Common Device Name
PASSER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
MDR Report Key10311567
Report Number1219602-2020-01115
Device Sequence Number19038765
Product Code HWQ
UDI-Device Identifier03596010374691
UDI-Public03596010374691
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number014396
Device Catalogue Number014396
Device Lot Number0238833
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 07/02/2020
Supplement Date Received by Manufacturer08/18/2020
Initial Report FDA Received Date07/22/2020
Supplement Report FDA Received Date08/20/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-