• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PULSION MEDICAL SYSTEMS SE PICCO CATHETER; PROBE, THERMODILUTION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PULSION MEDICAL SYSTEMS SE PICCO CATHETER; PROBE, THERMODILUTION Back to Search Results
Model Number PV2014L16N
Medical Device Problem Code Leak/Splash (1354)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/25/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Further information has been requested and investigation is ongoing.A supplemental emdr will be sent when the investigation is completed.Product not returned.
 
Event or Problem Description
Prior to puncture, the picco catheter was inspected.It was detected that the picco catheter connection to the pressure transducer connector end is leaking (less than 5 ml).The product has been replaced immediately.No harm or clinical consequences occurred.Manufacturer reference #: (b)(4).
 
Additional Manufacturer Narrative
The complained catheter has been returned for investigation.No indications for a material deficiency (like air bubbles, particles) could be detected by naked eye and under the microscope.A simulated leakage test of the complained device could confirm the leakage as described in the complaint.A dhr review could not identify any non-confomities or deviations relevant to the reported issue.A 100 % leakage test is performed during production.A systematic weakness in the production process is considered as unlikely due to the very low complaint rate (<0,01 % for all types of picco catheter).No similar complaints have been received for this batch.Considering the complaint investigation results and the appearance of the crack (linear shaped propagation, originating from luer lock connection) it is seen most probably that excessive force (potentially by overriding the connection between pressure transducer and catheter) in combination with alcoholic disinfectant has been applied.According to the customer no alcoholic disinfectant has been used.It cannot be exluded that any unknown tool has been used to tighten the connection.Upon the event occurrence the device was involved, but not used on a patient.The issue has been detected prior to puncture, when the catheter was inspected and connected to the pressure transducer.The evaluation results did not indicate that the device failed to meet its specification.The ifu states: ¿the luer-lock connection is to be tightened manually.The use of any tools or excessive force may damage the connection component and lead to leakiness.¿ and "usage of organic solvents for cleaning and disinfection may damage the catheter and lead to leakiness.¿ the issue will be further monitored on the market in order to indentify any trend.
 
Event or Problem Description
Manufacturer reference #: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PICCO CATHETER
Common Device Name
PROBE, THERMODILUTION
Manufacturer (Section D)
PULSION MEDICAL SYSTEMS SE
hans-riedl-strasse 17
feldkirchen 85622
GM  85622
MDR Report Key10312549
Report Number3003263092-2020-00009
Device Sequence Number11337055
Product Code KRB
UDI-Device Identifier04250094500658
UDI-Public(01)04250094500658(17)220228(11)170314(10)619483
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K072364
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 08/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2022
Device Model NumberPV2014L16N
Device Catalogue Number6885052
Device Lot Number619483
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/04/2020
Initial Date Received by Manufacturer 07/13/2020
Supplement Date Received by Manufacturer08/13/2020
Initial Report FDA Received Date07/23/2020
Supplement Report FDA Received Date08/13/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age48 YR
Patient Weight58
-
-