| Catalog Number |
20130-01 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/29/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The device will not be returned for evaluation.Without the return of the sample a comprehensive failure investigation cannot be performed, and a cause cannot be determined.If additional information becomes available a supplemental report will be submitted.
|
| |
|
Event or Problem Description
|
|
The event involved a spinning spiros® closed male luer, red cap where the spiros was not locking on the mating device and a leak was noted during a syringe administration of doxorubicin.It was reported that the leak was contained in a soaker pad.There was no obvious damage noted on the product.There was no blood loss or bleed back and no patient harm was noted when the issue occurred.
|
| |
|
Additional Manufacturer Narrative
|
|
H10 - no product samples were returned for investigation; however, a series of photographs were returned showing the label of the 20130-01 spinning spiros with lot number 4749754.The actual product assembly was not shown.The lot number identified used of a molded poppet that molding anomaly on the sealing surface that can result in leakage.The probable cause of the leaking spiros is due to a molding anomaly on the sealing surface of the poppet sub-component.The device history review for lot 4749754 was reviewed and a poppet sub-component was found that a molding anomaly on the sealing surface that can lead to leakage.
|
| |
|