• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-551NAP; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. 530G INSULIN PUMP MMT-551NAP; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-551NAP
Medical Device Problem Code No Display/Image (1183)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/18/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event or Problem Description
Information received by medtronic indicated that the insulin pump had a blank display.Customer stated the battery was changed multiple times, but the issue was not resolved.No harm requiring medical intervention was reported.The insulin pump will not be returned for analysis.
 
Additional Manufacturer Narrative
Device monitored for several days and no blank display noted.Device received with normal operating current.Device passed displacement test and self test.Device passed off no power test.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
530G INSULIN PUMP MMT-551NAP
Common Device Name
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key10313100
Report Number2032227-2020-151192
Device Sequence Number2613161
Product Code OZO
UDI-Device Identifier00643169503656
UDI-Public(01)00643169503656
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-551NAP
Device Catalogue NumberMMT-551NAP
Device Lot NumberA4551NAPJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/29/2020
Initial Date Received by Manufacturer 07/18/2020
Supplement Date Received by Manufacturer08/18/2020
Initial Report FDA Received Date07/23/2020
Supplement Report FDA Received Date08/19/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age36 YR
Patient Weight64
-
-