MAUDE Adverse Event Report: BARD ACCESS SYSTEMS POWERLOC MAX POWER-INJECTIBLE INFUSION SET 20G X 1.0 IN; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BARD ACCESS SYSTEMS POWERLOC MAX POWER-INJECTIBLE INFUSION SET 20G X 1.0 IN; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Model Number N/A
Device Problems Material Deformation (2976); Activation Problem (4042)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/24/2020
Event Type  malfunction  
Manufacturer Narrative
The manufacturer has received the sample and will evaluate.Results are expected soon.A lot history review (lhr) of asdrs0197 showed no other similar product complaint(s) from this lot number.
 
Event Description
It was reported that a needle would not lock to make it a safe sharp after use, it was further reported that the needle was bent.
 
Manufacturer Narrative
The information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, trend analysis, applicable previous investigation(s), sample (if available), and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of inability to activate the safety mechanism was confirmed and the cause was determined to be use related.The product returned for evaluation was a one 20ga x 1" powerloc max safety infusion set.The returned product sample was evaluated and the needle was observed to be bent at multiple points along the length of the needle shaft.The following observations were noted during the sample evaluation: usage residue was seen on the sample which proved that the product had experienced at least some use.The bend in the needle had occurred away from the needle base which can be caused by device manipulation during/following port access.The needle tip was barbed suggesting contact between the needle and port base a functional test of the safety mechanism found that the safety mechanism would not engage over the needle tip due to the bend within the needle.Force applied at an angle to the needle axis, or if the needle is not inserted perpendicular into the port septum, can lead to bending of the needle shaft.It is advised to verify that the needle length is correct based on port reservoir depth, tissue thickness and the thickness of any dressing beneath the bend of the needle; if too long, needle and/or port may be damaged at insertion.Additionally, avoid excessive manipulation once the needle is in the port.No potential damage related to the manufacturing process was noted on the complaint sample.A lot history review (lhr) of asdrs0197 showed no other similar product complaint(s) from this lot number.
 
Event Description
It was reported that a needle would not lock to make it a safe sharp after use, it was further reported that the needle was bent.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWERLOC MAX POWER-INJECTIBLE INFUSION SET 20G X 1.0 IN
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
BARD ACCESS SYSTEMS
605 n. 5600 w.
salt lake city UT 84116
MDR Report Key10313764
MDR Text Key200042926
Report Number3006260740-2020-02609
Device Sequence Number1
Product Code FPA
UDI-Device Identifier00801741047442
UDI-Public(01)00801741047442
Combination Product (y/n)N
PMA/PMN Number
K153440
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 08/19/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number0142010
Device Lot NumberASDRS0197
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/14/2020
Event Location Hospital
Date Manufacturer Received07/30/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-