| Model Number |
MMT-751LNAS |
| Medical Device Problem Codes |
Mechanical Problem (1384); Device Difficult to Program or Calibrate (1496)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/17/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump had keypad anomaly.Customer stated the insulin pump's act button was not registering properly when pressing down and had to press twice and slower to rewind.No harm requiring medical intervention was reported.The insulin pump will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with intermittent button response due to flattened act button dome switch.No button error alarm during testing.No unlocked lcd keypad connector.Device passed the displacement test and rewind test.No unexpected rewind anomalies noted.(b)(4).
|
| |
|